Difference between revisions of "Software Configuration"

From Click Wiki
Jump to: navigation, search
Line 30: Line 30:
 
In order to proceed, the calibration kit .xml file must be loaded, by clicking on the button'''Load Calibration Kit from File'''.
 
In order to proceed, the calibration kit .xml file must be loaded, by clicking on the button'''Load Calibration Kit from File'''.
  
[[File:LoadCal.PNG]]
+
[[File:LoadCal.png]]
  
 
This .xml file can be found in the scanner's internal memory, identified by the computer as if it were a Usb flash drive.
 
This .xml file can be found in the scanner's internal memory, identified by the computer as if it were a Usb flash drive.
  
[[File:CalKitLocation.PNG]]
+
[[File:CalKitLocation.png]]
  
 
'''NOTE: CHECK THAT THE NUMBER OF THE FILE AND THE NUMBER OF THE PLATE YOU HAVE IN THE ACCESSORIES BOX COINCIDE. IF THEY DO NOT CONTACT YOUR DEALER BEFORE CALIBRATING'''
 
'''NOTE: CHECK THAT THE NUMBER OF THE FILE AND THE NUMBER OF THE PLATE YOU HAVE IN THE ACCESSORIES BOX COINCIDE. IF THEY DO NOT CONTACT YOUR DEALER BEFORE CALIBRATING'''
  
Insert then the plate in the scanner and click the button '''Head Calibration'''
+
To insert the calibration plate in the scanner, detach the black support on the back of the plate simply by pulling, as it is held by magnets.
  
[[File:MasterinScanner.PNG]]
+
[[:File:Master-click.jpg]]
  
[[File:HeadCal.PMG]]
+
Release the two screws of the impression holder and insert the master support, tightening the screws as much as possible to fix it properly.
 +
 
 +
Place the calibration plate in the scanner taking care in positioning it keeping the part with one bold dot on the left and the side with two bold dots on the right.
 +
 
 +
[[File:MasterinScanner.png]]
 +
 
 +
Click the button '''Head Calibration''' to start calibration.
  
 
===Movement System Calibration===
 
===Movement System Calibration===

Revision as of 10:21, 31 May 2017


With the first launch of the software after installation comes the configuration of the 1Click software. This first configuration is fully guided and will allow the user to be operative scanning and sending files in just a couple of minutes.

Setting the Configuration File

The first step of the process allows to set the configuration file of the Scanner. The Configuration file holds all the information on the various components of the user's specific scanner, therefore it is crucial for the functioning of the system.

LoadConf.PNG

To set it, make sure that the scanner is correctly connected and switched on and click on the Get From Scanner button to load it automatically.

LoadedConf.PNG

Once the file has been succesfully loaded, in the box will be shown:

  • the scanner's serial number
  • the scanner's components

It is also possible to load a configuration from and external file, using the button Load From File, though this kind of operation will be performed in very rare cases (for instance if Open Technologies had to modify the configuration file of the scanner and the internal file is no longer valid).

Click next in the dialog to follow with the configuration, click Cancel to abort.

System Calibration

The following step is the System Calibration starting with the Optical Head and then calibrating the movement system.

Head Calibration

In order to proceed, the calibration kit .xml file must be loaded, by clicking on the buttonLoad Calibration Kit from File.

File:LoadCal.png

This .xml file can be found in the scanner's internal memory, identified by the computer as if it were a Usb flash drive.

File:CalKitLocation.png

NOTE: CHECK THAT THE NUMBER OF THE FILE AND THE NUMBER OF THE PLATE YOU HAVE IN THE ACCESSORIES BOX COINCIDE. IF THEY DO NOT CONTACT YOUR DEALER BEFORE CALIBRATING

To insert the calibration plate in the scanner, detach the black support on the back of the plate simply by pulling, as it is held by magnets.

File:Master-click.jpg

Release the two screws of the impression holder and insert the master support, tightening the screws as much as possible to fix it properly.

Place the calibration plate in the scanner taking care in positioning it keeping the part with one bold dot on the left and the side with two bold dots on the right.

MasterinScanner.png

Click the button Head Calibration to start calibration.

Movement System Calibration

To follow with the movement system calibration insert an impression in the scanner as shown in the image and click on the button Movement Calibration

File:ImpressioninScanner.PNG

MovCal.PNG

Click next in the dialog to follow with the configuration, click Cancel to abort.

Set the Trimming

In this step the user can configure an automatic trimming that will automatically erase all the information located under the scan height plane.

File:Trimming:PNG

Move the slider on the bar in the dialog to modify the scan height, taking particular care in not colliding with essential parts of the impression.

Trimmed.PNG

Click next in the dialog to end the first configuration procedure.

Dental 2 5 wiz prev.png 120px-Homebutton.png Dental 2 5 wiz next.png