First Software Configuration

From Click Wiki
Jump to: navigation, search


With the first launch of the software after installation comes the configuration of the 1Click software. This first configuration is fully guided and will allow the user to be operative scanning and sending files in just a couple of minutes.

Setting the Configuration File

The first step of the process allows to set the configuration file of the Scanner. The Configuration file holds all the information on the various components of the user's specific scanner, therefore it is crucial for the functioning of the system.

LoadConf.PNG

To set it, make sure that the scanner is correctly connected and switched on and click on the Get From Scanner button to load it automatically.

LoadedConf.PNG

Once the file has been succesfully loaded, in the box will be shown:

  • the scanner's serial number
  • the scanner's components

It is also possible to load a configuration from and external file, using the button Load From File, though this kind of operation will be performed in very rare cases (for instance if Open Technologies had to modify the configuration file of the scanner and the internal file is no longer valid).

Click next in the dialog to follow with the configuration, click Cancel to abort.

System Calibration

The following step is the System Calibration starting with the Optical Head and then calibrating the movement system.

Head Calibration

In order to proceed, the calibration kit .xml file must be loaded, by clicking on the buttonLoad Calibration Kit from File.

LoadCal.PNG

This .xml file can be found in the scanner's internal memory, identified by the computer as if it were a Usb flash drive.

CalKitLocation.PNG

NOTE: CHECK THAT THE NUMBER OF THE FILE AND THE NUMBER OF THE PLATE YOU HAVE IN THE ACCESSORIES BOX COINCIDE. IF THEY DO NOT CONTACT YOUR DEALER BEFORE CALIBRATING

To insert the calibration plate in the scanner, detach the black support on the back of the plate simply by pulling, as it is held by magnets.

Master-click.jpg

Release the two screws of the impression holder and insert the master support, tightening the screws as much as possible to fix it properly.

Supporto-master.png

Place the calibration plate in the scanner taking care in positioning it keeping the part with one bold dot on the left and the side with two bold dots on the right.

MasterinScanner.png

Click the button Head Calibration to start calibration.

Movement System Calibration

To follow with the movement system calibration insert an impression in the scanner as shown in the image and click on the button Movement Calibration. Remember that if you think of using the Triple Impression Module it is best to calibrate using the triple impression holder, so to have more precise plane cut settings.

MovCal.PNG

Click next in the dialog to follow with the configuration, click Cancel to abort.

Set the Trimming

In this step the user can configure an automatic trimming that will automatically erase all the information located under the scan height plane for the individual impression and the planes to separate the elements for the triple impression holder.

Move the slider on the bar in the dialog to modify the scan height, taking particular care in not colliding with essential parts of the impression.

Trimming.png

Then, if the Triple Impression Module is enabled in your license, click on the second tab Triple and move the sliders to establish the cut planes to separate the elements (upper from lower and individual from bite)

TripleTrimming.png

Click next in the dialog to end the first configuration procedure.

Access Calibration after first Configuration

Should it be needed, it is possible to repeat the calibration even after the first configuration of the system simply by accessing the options in the welcome page using the Option Button Optionbutton.PNG.
In the first Option Page, General, simply click Restart Scanner Configuration to restart the calibration procedure.

RestartConf.png